Abstract
Background: Right ventricular (RV) function is an important aspect of anesthesia management during orthotopic liver transplantation (OLT). Because of its geometrical complexity, assessment of RV dimensions with transesophageal echocardiography (TEE) is a difficult task. The aim of this prospective single-site study was to investigate the feasibility of intraoperative assessment of RV parameters based on reconstructive three-dimensional (3D) TEE and to compare the measurements to thermodilution-derived values acquired with a modified pulmonary artery catheter.
Material/Methods: Measurements were performed at four different time points during 30 OLT with 3D-TEE. At the same time comparative values of RV parameters were acquired with a fast-response thermistor pulmonary artery catheter.
Results: 3D reconstruction was feasible in all patients. RV dimensions measured with 3D-TEE averaged 119.4ml (±38.5 ml) for enddiastolic and 68.9 ml (±27.7 ml) for endsystolic volumes. The RV ejection fraction was 42.2% (±9.3%). The volumes obtained by thermodilution were 263.7 ml (±64.5 ml) enddiastolic and 159.3 ml (±47.5 ml) endsystolic, both significantly greater than by 3D-TEE, and the ejection fraction was found to be 39.5% (±8.4%). No correlation was found between the volumes or the function determined by either method.
Conclusions: Reconstructive 3D-TEE is a viable technique during OLT and leads to plausible RV parameters. However, no correlation was found with simultaneous measurements or parameters performed with thermodilution. However, based on our data cardiac output measurements by thermodilution appear reasonable. Due to both lack of agreement with 3D-TEE and extraordinary high RV volumes the question about the most valuable monitoring technique of RV dimensions and function during OLT can not finally be answered.
Keywords: Pulmonary artery catheter, transesophageal echocardiography, three-dimensional reconstruction, right ventricule, Orthotopic Liver Transplantation
In Press
18 Mar 2024 : Original article
Does Antibiotic Use Increase the Risk of Post-Transplantation Diabetes Mellitus? A Retrospective Study of R...Ann Transplant In Press; DOI: 10.12659/AOT.943282
20 Mar 2024 : Original article
Transplant Nephrectomy: A Comparative Study of Timing and Techniques in a Single InstitutionAnn Transplant In Press; DOI: 10.12659/AOT.942252
28 Mar 2024 : Original article
Association Between FEV₁ Decline Rate and Mortality in Long-Term Follow-Up of a 21-Patient Pilot Clinical T...Ann Transplant In Press; DOI: 10.12659/AOT.942823
02 Apr 2024 : Original article
Liver Transplantation from Brain-Dead Donors with Hepatitis B or C in South Korea: A 2014-2020 Korean Organ...Ann Transplant In Press; DOI: 10.12659/AOT.943588
Most Viewed Current Articles
05 Apr 2022 : Original article
Impact of Statins on Hepatocellular Carcinoma Recurrence After Living-Donor Liver TransplantationDOI :10.12659/AOT.935604
Ann Transplant 2022; 27:e935604
12 Jan 2022 : Original article
Risk Factors for Developing BK Virus-Associated Nephropathy: A Single-Center Retrospective Cohort Study of ...DOI :10.12659/AOT.934738
Ann Transplant 2022; 27:e934738
22 Nov 2022 : Original article
Long-Term Effects of Everolimus-Facilitated Tacrolimus Reduction in Living-Donor Liver Transplant Recipient...DOI :10.12659/AOT.937988
Ann Transplant 2022; 27:e937988
15 Mar 2022 : Case report
Combined Liver, Pancreas-Duodenum, and Kidney Transplantation for Patients with Hepatitis B Cirrhosis, Urem...DOI :10.12659/AOT.935860
Ann Transplant 2022; 27:e935860